Obstructive apnea
When the patient stops breathing the device no longer detects an effort, so spontaneous triggering does not occur and the device enters an apnea event. A 90% drop in airway flow for 10 seconds is treated as a possible obstructive apnea, and the timed backup cycle is activated (inspiration time 1 s, backup rate 15 bpm).
After 10 seconds the timed breaths are compared with the earlier spontaneous breaths. If the instantaneous minute volume has fallen below 80% of the target, this is classified as obstructive apnea. The device responds by raising EPAP in 0.5 mbar steps every five seconds — splinting the airway open — while pressure support is increased to maintain ventilation.